tinb21e1 (2016-06)
29
SP1 OFF and SP2 OFF
[Factory default = 200 mTORR]
These setpoints correspond to the pressures at which the relays
will turn off (de-energize). The relays will turn off when the pres-
sure is above the programmed pressure value. If you are unable
to decrease the values of SP1 OFF or SP2 OFF, you must first
go to SP1 ON or SP2 ON and decrease those values to a num-
ber lower than the values of SP1 OFF or SP2 OFF you are trying
to set.
RS485 ADDR
[Factory default = 1]
This is the lower nibble of the one byte RS485 device address.
Assuming the address offset (RS485 OFFSET) is equal to 0,
setting the ADDR to a 5 will make the address be 0x05 in hexa-
decimal. A 15 will set the ADDR to 0x0F in hexadecimal. Note
that the address (ADDR) must be used even when sending
RS232 commands.
RS485 OFFSET
[Factory default = 0]
This is the upper nibble of the one byte RS485 address. Assum-
ing the address (ADDR) is 0, setting the address offset (RS485
OFFSET) to a 5 will make the address be 0x50 hexadecimal.
Setting the address offset to 15 will make the device address be
0xF0 hexadecimal.
┌
----------------BINARY ADDRESS-----------
┐
ADDRESS
DECIMAL
┌
-----------------------ONE BYTE-------------------
┐
ADDRESS
HEXA-
DECIMAL
(BINARY)
ADDR OFFSET
┌
Upper nibble
┐
ADDR
┌
Lower nibble
┐
1
0 0 0 0
0 0 0 1
01
5
0 0 0 0
0 1 0 1
05
15
0 0 0 0
1 1 1 1
0F
16
0 0 0 1
0 0 0 0
F0
Summary of Contents for PGE500
Page 83: ...tinb21e1 2016 06 83 Notes...