Industrex M43ic / M43icN
page 29
01/2013 TK
Version 2.7 is equipped with a sub-program to control the filter unit of the developer:
Press the bottum
3
, use
6
to move the courser under "Options", press
7
, you
will see "Standby Refill Filter" use again
6
to move the courser under "Filter",
press
7
you will see:
The Area menu
sets the amount of square meters before filter alert.
The value is 0000-9999 . A value of 0000 turns off the filter alert.
Filter alert
after xxxx sqm
Save Cancel
The Log menu
shows the total area processed so far. This value is stored in the NVRAM.
The value is updated each time a replenishment is initiated. It depends on the „Replenish
after“ variable from the setup.
0000.00
Exit Clear
For example if Replenish after = 0.125sqm , the total processed area will be increased by
0.125 sq.m. at each replenishment.
At each replenishment the total processed area is checked for exceeding the alert value.
When the value has been exceeded, an error message appears „Change filter“.
After changing the filter, the user must go to programming mode menu Options/Filter/
Log and clear the processed area, so the area will be counted from 0 for the new filter.
5.12 FILTER CONTROL:
Area Log
WARNING: Call for qualified service personnel. The filter
may be exchanged by qualified service personnel only.
Summary of Contents for M43ic
Page 1: ...INSTRUCTION MANUAL 04 2011 TK NDT FILM PROCESSOR Industrex M43ic M43icN AT800 v2 8r09p and up ...
Page 3: ...Input speed DEV time Cycle time KODAKIndustrexM43ic Industrex M43ic M43icN ...
Page 49: ...page 43 01 2013 TK Industrex M43ic M43icN NOTES ...
Page 50: ...page 44 01 2013 TK Industrex M43ic M43icN NOTES ...
Page 51: ...page 45 01 2013 TK Industrex M43ic M43icN NOTES ...