46
Error
Reason
Measure
Fault in pressurization on cuff
· Not connected with the cuff
· Wearing the cuff, too loosely
· Air is leaking from the cuff
· Check the cuff connection and how it's
worn.
· If the air leaks out from the cuff, replace
cuff with a new one.
The pressure in the cuff unexpectedly
changes during pressurization
· Moving severely
· The air tube is folded
· The cuff is released
· The air tube falls out
· Measure the blood pressure in a stable state
after checking the device connection and
the cuff.
The pressure in the cuff unexpectedly
changes during exhaustion
· Moving severely
· The air tube is folded
· The cuff is released
· The air tube falls out
The signal of a blood pressure is too weak
· Measured in thick clothes
· The pulse is weak by nature
· If it was measured on the clothes, take off
the clothes and measure at bare skin.
· Check how the Cuff is worn.
The pressure exceeds 300 mmHg
· The user carried out the manual
pressurization excessively
· Severely moves
·When manual pressure is applied, pressurize
the appropriate amount. (Pressurize 30 to
40 mmHg higher than the expected blood
pressure.)
· Measure the blood pressure in a stable state.
It is a failure due to malfunction
Please contact the InBody Customer Center
(+82-80-501-3939).
Low battery voltage
Charge the lithium-ion batteries or replace
the AA batteries with new ones.
Troubleshooting
A. Error Message
If a problem is occurred, is displayed in the systolic blood pressure filed and
a letter or number appears in the diastolic blood pressure field.