26
∙ Month / Day Hour: Minute is displayed only when date and time are set.
∙ For more information on how to set the date and time, refer to “F. User Settings”.
∙ AUSCULT mode does not automatically measure blood pressure.
∙ AUSCULT mode does not detect heartbeats.
∙ In AUSCULT mode, pulse is not displayed.
∙ In AUSCULT mode, motion and irregular pulse wave are not detected.
∙ If no marking is performed, the measurement is automatically terminated at a certain
pressure or less.
Note
∙ If pressure exceeds 300 mmHg, the measurement is terminated and an error is displayed.
∙ Use the manual pressure function while checking the patient’s condition. Long term
pressurization may cause pains.
Note
- The AUSCULT mode allows manual pressurization to be performed at the desired
pressure during exhaust in case the patient cannot hear.
❶
Press and hold the [▲] button during exhaust.
❷
The manual pressure indicator is displayed while pressing this button, and pressurization begins.
❸
Release the [▲] button after the appropriate pressurization, and the pressure stops.
- If rapid exhaust is required, manual exhaust is possible to achieve desired pressure.
❶
Press and hold the [▼] button during exhaust.
❷
The manual pressure indicator is displayed while pressing this button, and rapid exhaust begins.
❸
Release the [▼] button after the appropriate pressurization, and the rapid exhaust stops
and switch to normal exhaust speed.
During Exhaust
Press and Hold to Start
Manual Pressurization.
During Exhaust
Press and Hold to Start
Rapid Exhaust.