52
E. EMC Declaration
This EUT is intended for use in the electromagnetic environment specified below. The customer
or the user of the EUT assure that it is used in such an environment.
Immunity Test
IEC 60601-1-2
Test Level
Compliance Level Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
100 kHz
repetition frequency
± 2 kV
100 kHz
repetition frequency
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge Line-to-line
IEC 61000-4-5
± 0.5 kV, ± 1 kV
± 1 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge Line-to-
ground
IEC 61000-4-5
± 0.5 kV, ± 1 kV,
± 2 kV
± 2 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips
IEC 61000-4-11
0 % UT; 0.5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % UT; 0.5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the EUT
image intensifier requires continued
operation during power mains
interruptions, it is recommended that
the EUT image intensifier be powered
from an uninterruptible power supply
or a battery.
0 % UT; 1 cycle
and 70 % UT; 25/30
cycles
single phase: at 0°
0 % UT; 1 cycle
and 70 % UT; 25/30
cycles
single phase: at 0°
Voltage
interruptions
IEC 61000-4-11
0 % UT; 250/300
cycle
0 % UT; 250/300
cycle
RATED power
frequency
magnetic fields
(50/60Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Summary of Contents for BPBIO250
Page 1: ...BPBIO250 User s Manual ...
Page 4: ......
Page 54: ......
Page 55: ......