68
Level
trOll
Operator’s Manual
0052210 rev. 006
When the nose
cone is
removed, the
sensitive pressure sensor
diaphragm is completely
exposed. Do not touch this
area with any object!
Replace the nose cone as
soon as possible.
seCtion 9: Care & maintenanCe
StoraGe
Store the Level TROLL clean and dry. Place the protective red dust cap on
the cable end, or store with cable attached to protect the connector pins
and o-ring.
Store the instrument where it will be safe from mechanical shocks that
may occur, such as rolling off a bench onto a hard surface.
Protect the instrument from temperature extremes. Store within a
temperature range of -40°C to +80°C (-40°F to +176°F).
GeneraL maIntenanCe
Cleaning—body and Front end
Clean the Level TROLL body with water and a soft brush or plastic
scouring pad, or soak overnight in a mild acidic solution, such as
household vinegar.
If the ports in the front end are clogged with silt or mud, try the following:
• Swish the instrument vigorously in a bucket of clean water
• Apply a gentle squeeze of water from a wash bottle
• In severe cases, remove the nose cone and clean out the holes with a
soft brush or pipe cleaner
To avoid damage to the pressure sensor diaphragm,
do not insert any
object into the sensor opening or attempt to dig out dirt or other materials.
If contamination cannot be removed using the recommendations above,
please contact In-Situ Inc. for cleaning.
Damage caused by digging or scraping in the pressure sensor
opening to remove silt, mud, etc. is not covered by the warranty.
Nose cone removed
Nose cone in place