69
Level
trOll
Operator’s Manual
0052210 rev. 006
Do not
submerge
the cable
connector; do not immerse
in any fluid.
tWist-loCk ConneCtors
Keep the pins on all connectors free of dirt and moisture by using the soft
protective dust cap when cable is not attached.
Cable vent tube (vented Cable)
Vented cable assures that atmospheric pressure is the reference pressure
to the vented pressure sensor diaphragm.
The vent tube should not be
blocked, kinked, or otherwise obstructed.
Such obstructions will cause
barometric pressure to appear in measurements, and errors will be
introduced due to thermal expansion and contraction of air within the vent
tube and probe body.
The recommended minimum bend radius is 13.5 mm (0.54 in), which is
twice the cable diameter.
batterIeS
Internal batteries in the Level TROLL are not user-replaceable.
The
approximate percentage remaining is displayed on the Dashboard when
the Level TROLL is connected in software.
seCtion 9: Care & maintenanCe
The minimum
bend radius for
vented cable is
13.5 mm (0.54 in).
TIP:
If batteries
are completely
exhausted,
remember that external
power and battery pack
options are available.
TIP:
Be sure
to replace the
desiccant when
it appears pink. Expired
desiccant can allow
water build up in the vent
tube, causing a blockage
resulting in inaccurate data.