Error messages and Trouble shooting
8.6
Error messages
Error text in display
Explanation
Solution
If this icon is
displayed…
…exit
the application before
removing the USB
memory stick.
USB flash drive was removed before
leaving the method
The saved file may be damaged,
because it was not closed correctly
To avoid data loss please exit the method first
before disconnecting the USB flash drive
Calibration failure
(UV on Reference
channel)
Submicroliter cell or cuvette in the
cell holder by switching on the
instrument.
Instrument was too cold.
Remove cuvette or submicroliter cell from the
cell holder, turn off the instrument and start
again.
Did you start the NanoPhotometer
®
P-Class
directly after delivery? If yes, turn it of and wait
30 min before switching the unit on.
If both suggestions doesn’t help disconnect
the power completely for at least 10 seconds
and then reconnect and restart the system
again.
Cell holder obstructed
Submicroliter cell or cuvette in the
cell holder by switching on the
instrument.
Remove cuvette or submicroliter cell from the
cell holder, turn off the instrument and start
again.
No light on Reference
channel
Submicroliter cell or cuvette in the
cell holder by switching on the
instrument.
Cell holder has been removed from
the instrument and placed back in
the wrong position.
Remove cuvette or submicroliter cell from the
cell holder, turn off the instrument and start
again.
Remove the cell holder and place back in the
right position.
If both suggestions doesn’t help disconnect
the power completely for at least 10 seconds
and then reconnect and restart the system
again.
Calibration problem
(UV/IR) on Reference
channel
Low light levels.
Disconnect the power for quite some time to
make sure that the problem is not occurring
due to low energy sent to the Xenon flash
lamp.
Waiting for software
update (Keyboard)
The instrument has been turned on
and off too fast.
Disconnect the power completely for at least
10 seconds and then reconnect and restart
the system again.
Calibration problem:
possible lamp failure
Insufficient provision of electricity.
Power supply does not deliver 18V.
Check all connections if the cables etc. are
sticking right.
Check correct power supply 18 V is attached.
If possible try another power supply.
Please contact the Implen Support Team ([email protected], Phone +49-89-7263718 20) if none of the mentioned
solution helps to solve the problem or if another error message should appear on the NanoPhotometer
®
P-Class
display.