- 36 -
S-2018
REV. D
NIOSH Performance Criteria for the
Sentinel XL™ PAPR CBRN
System Level Agent Challenge (Smart Man)
Agent
Challenge Concentration Duration of Challenge (minutes)
Mustard (HD) Vapor
50mg/m
3
30
Sarin (GB) Vapor
210 mg/m
3
30
Chemical Test Agents (Filter Gas Life)
Test Agent
Challenge
Concentration (ppm)
(170 lpm @ both
25% and 80% RH)
Allowable Breakthrough (PPM) Service Time (Min)
Ammonia (NH
3
)
1250
12.5
>30
Cyanogen Chloride (CK)
150
2
>30
Cyclohexane (C
6
H
12
)
1300
10
>30
Formaldehyde (HCHO)
250
1
>30
Hydrogen Cyanide (HCN)
470
4.7*
>30
Hydrogen Sulfide (H
2
S)
500
5.0
>30
Nitrogen Dioxide (NO
2
)
100
1 ppm NO
2
or 25 ppm NO **
>30
Phosgene (COCl
2
)
125
1.25
>30
Phosphine (PH
3
)
150
0.3
>30
Sulfur Dioxide (SO
2
)
750
5
>30
* Sum of HCN and C
2
N
2
**Nitrogen Dioxide breakthrough is monitored for both NO
2
and NO. The Breakthrough is determined by
which quantity, NO
2
or NO, reaches breakthrough first.
CUSTOMER SERVICE INFORMATION
If you need additional information on the
Sentinel XL™ PAPR CBRN
or any ILC Dover product, consult your
distributor or contact:
Customer Service
ILC Dover
One Moonwalker Road
Frederica, DE 19946-2080
Phone: 800.631.9567 OR 302.335.3911, ext. 506
Fax: 302.335.1320
Summary of Contents for S-5000 Series