23
SETTING
1. Languages:
The “
Language
“ menu option allows you to select the
desired language�
2. Display:
The ”
Display
“ menu option allows you to change the
background color and brightness of the working screen�
You can also activate the “Firmware Update Info“ screen
in this menu�
3. Sound:
The ”
Sound
” menu option allows you to activate/deacti-
vate the key tone�
4. Factory Settings:
Select the ”Factory settings” menu option by turning
and pressing the rotary/push knob� The system requests
confirmation to restore the factory settings� Pressing the
”OK” button resets all the system settings to the origi-
nal standard values set at dispatch from the factory (see
“Menu structure”)�
5. Communication:
The ”Device name” menu option allows you to edit the de-
vice name� This can relatively identify the paired
station
�
The ”
Bluetooth
®
” menu option allows you to activate/
deactivate the ”
Bluetooth
®
” function� A check mark
shows that the option is activated�
6. Information:
The ”
Information
” menu option offers you an overview
of the most important system settings of the device�
SAFETY
1. Time out:
1) Set:
In the menu option “Set”,you can determine a time limit
in the event of a communication breakdown between the
station
and the
WiCo
� The
station
continues to work
with the preset target values until the preset time value
has expired� Following that, the
station
runs using the
preset safety temperature and safety speed�
Note:
The initial time out is 30 seconds and you can de-
fine up to 60 minutes for this time limit�
2) Speed:
In the “Speed“ menu option, you can specify a appropri-
ate and safe speed for specified temperature�
Note:
The factory setting of the safe speed is 2000 rpm
and is activated after the time limit is set (see “Set”)�
3) Temperature:
In the “Temperature“ menu option, you can specify a
temperature that is appropriate and safe for the circulat-
ing task�
OUTGAS
In the menu option, you are allowed to set the outgas
speed from 2000 rpm to 4000 rpm and outgas interval
from 10 seconds to 240 seconds�
This function can be used when filling external devices
such as laboratory reactors�
Temperature (int)
Temperature-dependent control of the M2-switch�
By entering the boundaries of “Higher” and “Low”, the
temperature range is set in which the M2-switch can be
opened�
If you activate a rule “Int / ext> Set Temperature” is an
automatic control of the M2-switch dependence on the
target temperature�
By the “Hysteresis” value (set temperature hysteresis) the
control stability can be optimized�
Note:
int / ext stands according to the selected control
mode (internal (int) or external (ext)) for the actual tem-
perature to be controlled�
3. M3-Input:
Ext. Standby:
External standby input to stop the device functions Tem-
per and Pump�
ON:
Activate the ext� standby function� Device functions
are stopped in the event of an inverted initial state (ON)
at the input�
Default Status: Define the default status (OFF) of the input
as “Open” (high level) or “Closed” (low level)�
Note:
The initial safe temperature is 30 ºC and is acti-
vated after the time limit is set (see “Set”)�
2. Password:
In the “Password” menu, the menu settings can be locked
by a 3-digit password�
3. Safe Temp. Confirmation:
This menu allow you to confirm the safe temperature of
the
station
at start-up� This menu can confirm (safety
temperature of the
station
) at start-up of the set “Safe
Temp”� A check mark (√) indicates that the function is
activated�