Setting the safety temperature

Setting the safety temperature with screwdriver delivered 
with the device� 

The safety temperature setting will appear on the display�

Factory setting: maximum value�
Adjustment range: 0 to 110 °C�

Warning: 

The safe temperature limit must always be set to 

at least 25 ºC lower than the flash point of the fluid used� 

Commissioning

Note: 

Before commissioning, make sure that the device 

has not been moved in one hour!

Check whether the voltage given on the type plate corre-
sponds to the available power voltage� 

The socket used must be earthed 
(fitted with earth contact)� 

If these conditions have been met, the machine is ready for 
operation when the power plug is plugged in� 
If these conditions are not met, safe operation is not guaran-
teed and the machine could be damaged�

Observe  the ambient conditions (temperature, humidity, 
etc�) listed under “Technical Data”�

After pressing the power switch (

12

) of the 

station

, the 

screen of the 

WiCo

 displays the device designation and the 

software version after a beep� 

IKA

®

Circulator 

HRC Control

Version 1.2.008

Fig. 20

After several seconds, screen display the information of 

WiCo

� 

Information

Internal Max: 

85 °C

Internal Min: 

5 °C

External Max: 

85 °C

External Min: 

5 °C

Fluid: H

2

O (Water)

Fig. 21

Then the working screen appears and the device is ready 
for operation�

A

0.00

int °C

ext °C

int

5.00

2000

set °C

Pump rpm

Safe Temp °C

Fig. 22

Change the temperature setting with left knob (P)�
Change the pump speed setting with the right knob (R)�

In standby status, activate the tempering function by press-
ing the left knob (P)� The device start tempering function� 
Meanwhile, the pump starts to run at once� 
After pressing the power switch (12) the start-up of the 
compressor is delayed (about 3 minutes)�
In working status, press the right knob (R) to stop operation 
of the pump� The tempering function and the pump stops�

Hig

h

Low

Flui

d le

vel

IKA

®

 RC 2 

contr

ol

IKA

®

 HRC 2 

contr

ol

Fig. 19

+

-

A

0.00

int °C

ext °C

int

5.00

2000

set °C

Pump 

rpm

Safe T

emp °C

25.0

0

110.0

0

Summary of Contents for HRC 2 control

Page 1: ...Operating instructions EN 20000015725a IKA HRC 2 control HRC 2 control_112017...

Page 2: ...Designation 1 Wireless Controller WiCo 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 External temperature sensor socket 8 Multifunction port 9 Overflow 10 Backflow 11 Power...

Page 3: ...he WiCo 9 Mounting the WiCo to the station 9 Filling and draining 9 Fluid Standard information for IKA fluid 11 Moving the device 12 Charging the Battery Pack RB 1 rechargeable battery 12 Changing bat...

Page 4: ...rates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equip...

Page 5: ...inhibit fire or ex plosions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application sup port if you have any...

Page 6: ...union nuts caps of the pump outputs and inputs and blow compressed air through all the pump outputs and inputs Beware of the risk of burning due to delay in boiling Fluids Beware of the risk of burnin...

Page 7: ...below 20 C Only the rechargeable battery types recommended in the technical data may be used in the device Do not charge batteries that have leaked or that are discolored deformed or damaged in any ot...

Page 8: ...nection M 16 x 1 for IN and OUT directly or with the olives Screw the hose olives to the pump connection IN and OUT with union nuts Slide the hoses NW 12 onto the olives The hoses must be secured with...

Page 9: ...S IP 40 Model WiCo CBC 5 LPS IP 40 Note notch upward Mounting the WiCo to the station Note If the WiCo need be permanently attached to the station we strictly recommend to fasten the unlocking button...

Page 10: ...rain valve in counter clockwise direction with a straight screwdriver Fluid level information on the fluid level indicator Red Low Level Red High Level Fig 12 Note Pay attention to the fluid level inf...

Page 11: ...e Temperature 4 Viscosity Viscosity at 25 C for Heating Fluid HF Viscosity at 20 C for Cooling Fluid CF Viscosity at 25 C for Universal Fluid UF Dynamic viscosity mPa s is a product of kinematic visco...

Page 12: ...oss head clamp H 38 Holding rod IKA HRC 2control IKA HRC 2 control IKA HRC 2control Moving the device Empty all fluid in the bath before moving device from one place to other place The device can be l...

Page 13: ...evel U Menu button Press it once main menu is displayed Press it a second time back to the working screen V Lock button Lock Unlock the knobs and buttons Item Designation Function A On Off button Swit...

Page 14: ...ssing the power switch 12 of the station the screen of the WiCo displays the device designation and the software version after a beep IKA Circulator HRC Control Version 1 2 008 Fig 20 After several se...

Page 15: ...station the USB icon appears on the screen If the WiCo is not connected with station via a USB cable the data exchange between the station and the WiCo via Bluetooth In this case the Bluetooth icon i...

Page 16: ...This symbol indicates that the pump is activated Warning This symbol indicates that warning is active PC PC Control This symbol means that either the station or the WiCo is connected to a computer and...

Page 17: ...f you press the Menu button U the system skips directly back to the working screen If you press the Back button T the system skips back to the previous screen Menu Menu navigation Press the Menu butto...

Page 18: ...nment Set Temperature Actual Internal Temp Actual External Temp X Axis Y Axis 15 min 30 min 60 min 120 min Axis Scaling Maximum Minimum activated activated activated activated activated 85 C 5 C PROGR...

Page 19: ...be used in the event of special tem perature control requirements When Manual is selected the following parameters can be set for Internal int and External ext temperature control Kp Proportional coef...

Page 20: ...2 OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the measuring device 5 00 OC OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the meas...

Page 21: ...The scaling of the temperature axis Y axis is automati cally determined depending on the target temperature and the actual internal and external temperature The time axis X axis is permanently scaled...

Page 22: ...ult Status Define the default state OFF of the external valve as Open or Closed It depends on the valve type nor mally open or closed 3 Refill Use the external M1 valve to automatically check the leve...

Page 23: ...specified temperature Note The factory setting of the safe speed is 2000 rpm and is activated after the time limit is set see Set 3 Temperature In the Temperature menu option you can specify a tempera...

Page 24: ...ll duplex Character format character representation in accordance with data format in DIN 66 022 for start stop mode 1 start bit 7 character bits 1 parity bit even 1 stop bit Transmission speed 9600 b...

Page 25: ...ature control OUT_WD1 n Start the watchdog mode 1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the tempering and pump functions are switched off...

Page 26: ...y 0V only for Standby 7 reserved for later use do not connect 8 reserved for later use do not connect 1 2 3 4 5 6 7 8 Fig 35 Multifunction port USB micro B USB micro B USB micro A USB micro A USB cabl...

Page 27: ...e isopropyl alcohol Foodstuffs Water containing tenside Fuel Water containing tenside For materials which are not listed please request information from IKA application support Wear protective gloves...

Page 28: ...eating refrigerating off Heater switched off by safety circuit Check safety temperature set value fluid level Error 15 Pump off Heating refrigerating off Safety shutdown of the compressor Restart the...

Page 29: ...licone tube nominal width 12 H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12 FKM tube nominal width 12 mm Tubing insulations ISO...

Page 30: ...olution K 0 01 Classification according to DIN 12876 1 Class III FL suitable for flammable and non flammable fluids Safety circuit adjustable C 0 110 Safety temperature display TFT LCD on WiCo Heating...

Page 31: ...ient temperature C 5 40 Ambient humidity relative 80 IP code according to EN 60529 IP 40 Interface USB RB 1 Battery pack Voltage V 3 7 Charging capacity mAh 2000 Charging time h 4 5 Working time h 15...

Page 32: ...ble for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with the in stru...

Page 33: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Reviews: