![IKA HRC 2 control Operating Instructions Manual Download Page 21](http://html1.mh-extra.com/html/ika/hrc-2-control/hrc-2-control_operating-instructions-manual_624652021.webp)
21
PUMP:
Limits:
In “Limits“ menu option, you are allowed to set the max-
imum and minimum speed to the pump, and also the
maximum pressure�
PROGRAMS:
Under programs, 10 user-defined temperature-time pro-
files can be created� A program can consist of up to 10
segments�
Once a program has been selected, the following options
are available:
1. Start:
Starts the program upon request of loop mode�
1) Infinite loop:
Upon completion of the last segment,
the program continues with the first segment until you
end the program by stopping a device function�
2) Loop Count:
Indicates the total number of loop cycles
until program end�
Note:
At the end of the program all device functions are
switched off�
2. Edit:
Edit/change program�
1) Seg No.:
Segment number�
2) Ctrl. Sensor (int/ext):
Determines whether control is
through the internal (int) or external (ext) temperature
sensor�
3) Temp.:
Target temperature�
4) Ctrl.Mode (Time / +/- x.x K):
In "Ctrl�Mode Time" the
target values and settings of the segment are valid for the
duration indicated in the column "Time hh:mm"�
Afterwards, the next program segment is automatically
executed�
In "Ctrl�Mode +/- x�xx K", the hysteresis (tolerance) of the
actual temperature to the target temperature is set (e�g�
+/- 0�1 K)� The target values and settings of the segment
are valid until the actual temperature reaches the target
tempe/- hysteresis for the first time�
Afterwards, the next program segment is automatically
executed�
GRAPH:
In this menu, you can set the options for the time-tem-
perature diagram�
1. Automatic:
The scaling of the temperature axis (Y-axis) is automati-
cally determined, depending on the target temperature
and the actual internal and external temperature�
The time axis (X-axis) is permanently scaled to 30 minutes�
2. Manual:
1) Axis Assignment:
The temperature values to be displayed can be selected�
2) Axis scaling:
Scaling of the time (X) and temperature axis (Y) can be
selected or set�
DISPLAY:
1. Main Screen:
In the “Main Screen“ menu option, you can specify what
information will be displayed on the screen�
2. Device Number:
In the “Device number“ menu option, you are allowed to
edit the device number from 0 to 99 that will be shown
on the main screen�
When using multiple units, it may be helpful to identify
WiCo
and
station
�
Note:
Labeling device number to the
station
�
MODE:
1. Operating Mode A:
After power-on/power failure no automatic restart of
functions�
2. Operating Mode B:
After power-on/power failure automatic restart of func-
tions, depending on previous settings�
3. Operating mode C:
Set values (set in A or B) cannot be changed�
After power-on/power failure automatic restart of func-
tions, depending on previous settings�
4. Operating mode D:
Confirmation request for set value changes, if functions
are active� After power-on/power failure no automatic
restart of functions�
TIMER:
1. Set:
You can set a target time (duration) in this menu option�
When device functions are started normally, this time is
displayed on the working screen� The device functions
stop automatically once this time has expired� The run-
ning time is then displayed again on the display�
Note:
To deactivate the target time specification, set the
target time to 00:00�
2. Display:
Activate the timer display on the main display (working
screen)�