7.4.1 Installation in open tanks
►
For installation in open tanks, use a metal fixture to install the unit. It serves
as a launching plate (R); minimum size: 150 x 150 mm for a square fixture,
150 mm diameter for a circular fixture.
►
If possible, mount the unit in the middle of the fixture. The distance D2 must not
be below the indicated min. distances
.
D1
R
D2
L:
D2:
R:
Min. 150 mm.
Note the required min. distances
Launching plate