Note min. connection piece diameter D according to the figure / table below.
Select a connection piece height that is smaller than the connection piece
diameter.
• The rod must adhere to the following min. distances to the tank wall, objects
(B) in the tank and at the tank bottom:
A
1
B
A
2
A
3
D
Installation distances with adjustment
Installation distances without adjustment
A1: 10 mm
A1: 10 mm
A2: 20 mm
A2: 50 mm
A3: 20 mm
A3: 50 mm
D:
≥ DN 30
D: No connection piece allowed
• With intensely agitated medium (flow, agitator etc.) or heavy soiling:
►
Adhere to increased minimum distances to avoid that the rod gets into
contact with the tank wall or structures in the tank.