Select [cOP].
►
Press [SET] and keep it pressed.
> [cOP] flashes for some seconds; then the continuous display indicates
that the adjustment is being made.
> If the adjustment is successful, [donE] is displayed.
►
Confirm with [Mode/Enter].
> If the adjustment is not successful, [FAIL] is displayed.
►
If necessary, lower the level or correct the position of the overflow
prevention [OP] and repeat the adjustment operation.
cOP
Minimum distance between the overflow prevention OP and the level during
adjustment:
Table
10-1
[cm]
[inch]
LK3122
2.0
0.8
LK3123
3.5
1.4
LK3124
5.0
2.0
The position of the overflow protection OP can be determined by calling up
the parameter [OP]. Note the offset if necessary.
The current level is to be determined manually since the unit is not yet
ready for operation before the adjustment.
When the overflow prevention is activated ([OP] = [value ...]), an
adjustment [cOP] must be carried out each time:
• [MEdI] or [OP] was changed. In this case
appears in the display.
• The installation position (height, orientation) was changed.
• The connection between the sensor and the tank ground (e.g. cable
length) was changed.
With deactivated overflow prevention [MEdI] = [Auto] or [OP] = [OFF] it is
necessary for the unit for applying the basic settings and adaptation to the
medium and installation environment:
1. to be installed in the application
2. to be reinitialised.
►
Switch the operating voltage off and on again.