Manufacturer’s EMC guidelines
54
The Tel-O-Graph
®
BT has been tested with the following frequencies:
Measurement of
interference
emissions
Test level
Compliance level
RF interference
emissions according to
IEC 61000-4-3
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz Sinus)
PM; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz Sinus)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz