Error messages
44
Error
description
Possible cause
Remedy
Err 5
Power pack or battery voltage
too low.
Change the batteries (Chapter 5.2).
Battery contacts are corroded.
Clean the battery contacts with a cotton cloth and a little
alcohol.
Err 6
Air congestion
Check the blood pressure cuff for air congestion or a kink in
the air hose. If there is a kink in the air hose, unkink it,
otherwise send the blood pressure monitor to your
healthcare provider/physician or the manufacturer.
Blood pressure cuff incorrectly
connected.
Connect the blood pressure cuff to the blood pressure
monitor
(Chapter 6.2).
Leak in the blood pressure cuff
or air hose
Replace the blood pressure cuff with the air hose.
Err 9
Residual pressure in the blood
pressure cuff
Wait until the blood pressure cuff has deflated completely.
Err 10
Internal error
If this error occurs repeatedly, send the blood pressure
monitor to your healthcare provider/physician or directly to
the manufacturer for checking.
Abr. (Abort)
Measurement cancelled on
request
Do not press this button when measuring, unless you want
to abort the measuring.