background image

14

13

Line Frequency 

Within this menu selection, the ST-1THD displays the line frequency of the
current.  There is no submenu for this menu selection.

Recall Memory Menu (Event Recording)

Stored data can be viewed from the Recall Memory menu.   Pressing the
display advance button for three seconds enters the submenu.  The sub-
menu is identified by a blinking “rCL” on the LED display.  Once in the sub-
menu, pressing the display advance button gives access to the stored val-
ues from memory.  These values are accessed through the outlet circuit
testing menu.  While stored values are being displayed, the LED will blink
to differentiate these values from the "live" values being captured in stan-
dard testing mode.  Navigating through the outlet circuit testing menu while
in recall memory mode is identical to navigation through the menu in stan-
dard operation.

While the ST-1THD is in operation, the unit will analyze and store data on
the various measurements.  These values can be permanently stored into
the memory from the Recall Memory submenu.  Pressing the display
advance button for three seconds when a blinking “rCL” is shown on the
display will store all of the data into memory, overwriting any previous data
in memory.  

It is important to note that pressing the blinking “rCL” to exit viewing
recorded data stores new data into permanent memory.  If you wish to
maintain the stored data, but want to return to standard operation mode,
unplug the SureTest and re-insert it into the receptacle.  The SureTest will
return to operation mode.

• Menu Heading - “rCL”
• Measurement at Menu Heading - None
• Submenu - yes

Blinking “rCL” - Allows access to stored values from memory 
“Sto” - Stores values into memory

Testing with the #61-181 Clamp Adapter

These tests can only be accessed by using the #61-181 amp clamp adapter
with the model ST-1THDC.  Power must be supplied to the unit by an exten-
sion cord or battery pack to test the unit with the clamp adapter.

When the adapter is plugged into the ST-1THDC, the LED display will show
“Ct” indicating that the adapter is correctly plugged into the unit.  The ST-
1THDC will then automatically enter into the current probe programming
menu. The LED display cycles between the menu heading, and the measure-
ment every four seconds.  Press the display advance button for one second
to move to the menu.  To enter the submenu, press the display advance but-
ton for three seconds.  To navigate from measurement to measurement with-
in the submenu, press the display advance button.  While in the submenu,
pressing and holding the display advance button for three seconds brings
the unit back to the menu structure.

1. Insert the #61-181 adapter into the ST-1THDC
2. The LED display shows “Ct” indicating recognition of the adapter
3. The LED display cycles between “cA”, indicating that True RMS current

is being measured through the amp clamp adapter, and the current mea-
surement every four seconds.

4. Press and hold the display advance button for three seconds to access

the Current submenu.

5. The LED display alternates between “Hi”, indicating that the highest cur-

rent is being measured, and the measurement every four seconds

6. Press the display advance button for one second to move to the next test

within the Current submenu.

7. Press and hold the display advance button for three seconds to return to

the main menu.

Summary of Contents for SureTest ST-1THD

Page 1: ...nd Ground Conductors Min Max Avg Peak Neutral to Ground Voltage THD in the Line Voltage and Neutral Current THD per Harmonic up to the 31st Harmonic Line Frequency General Operation The SureTest Model...

Page 2: ...ground conductor Specifications Case Construction GE Cycolac plastic molded Operating Range All measurements except voltage drop FG drop FG conductor resistances 80 300 Volts AC 50 60 Hz Voltage drop...

Page 3: ...HD automatically enters into the out let circuit testing programming menu The LED display cycles between the menu heading and the measurement every four seconds Press and hold the display advance butt...

Page 4: ...edicated circuit Menu Heading Gn Measurement at Menu Heading True RMS Voltage Ground to Neutral Submenu yes Hi Highest Ground to Neutral Voltage Lo Lowest Ground to Neutral Voltage A Average Ground to...

Page 5: ...ecall Memory submenu Pressing the display advance button for three seconds when a blinking rCL is shown on the display will store all of the data into memory overwriting any previous data in memory It...

Page 6: ...a Sto Recording Completed Line Voltage THD Menu Menu Heading LT Measurement at Menu Heading Line Voltage THD Submenu yes HI Highest THD Lo Lowest THD A Average THD LC Crest factor of the line voltage...

Page 7: ...nt at Menu Heading None Submenu yes Blinking rCL Allows access to stored values from memory Sto Stores values into memory Testing with the 61 181 Clamp Adapter These tests can only be accessed by usin...

Page 8: ...on the adapter to a ground such as a metal water or gas pipe prior to GFCI testing 61 176 Isolated Ground Adapter This adapter allows the operator to verify that a receptacle is completely iso lated f...

Page 9: ...e voltage drops the problem is most likely in the circuit between the panel and the first receptacle or at the panel If all branches are in high resistance the problem may be in the cir cuit from the...