Appendix A. HART Command
The often HART universal commands are listed as below.
Command 0: Read Unique Identifier
Request data bytes: none
Response data bytes: 2+12 = 14
Index
Format
Description
Byte 0
uint8
Response code 1
Byte 1
uint8
Response code 2
Byte 2
uint8
254
Byte 3
uint8
Manufacturer ID
Byte 4
uint8
Manufacturer
’
s device ID
Byte 5
uint8
Number of preambles needed in the request
Byte 6
uint8
Command set revision number
Byte 7
uint8
Transmitter specific revision code
Byte 8
uint8
Software revision
Byte 9
uint8
Hardware revision
Byte 10
uint8
Flags
Byte 11~13
uint24
Device ID number (MSB first)
Command 1: Read Primary Variable
Request data bytes: none
Response data bytes: 2+5 = 7
Index
Format
Description
Byte 0
uint8
Response code 1
Byte 1
uint8
Response code 2
Byte 2
uint8
Unit code
Byte 3~6
float
Primary Variable
HRT-710 / HRT-310 User Manual
Version 1.31
Page
:
136