8-17-1-3-2 Adjust Regional Environmental Parameters
Users can change the regional environmental parameters as well as the global
parameters for individual thermal analysis objects during or after capture (see
Section 8-17-2 Global Environmental Settings
).
Adjust the regional parameters of an ROI as follows:
1.
Tap an existing temperature analysis object to select it. A Pop-up Menu
appears at the top of the Image Display Window.
2. Using the touch screen, tap the Regional Parameters icon:
3. Tap an input field next to an option: Emissivity, Reflected Temperature,
Distance, Atmospheric Transmittance, or Atmospheric Temperature.
4. Press-and-hold; then, drag up or down on the rotary dial, or tap - or +, to
adjust the setting.
5.
Tap ✓ to save changes. Tap X to cancel changes. The window closes.
6. Tap anywhere on the Image Display Window to close the Pop-up Menu.
Parameters can also be changed retroactively within the image gallery.
3.
4.
5.
REGIONAL PARAMETERS CAN DIFFER FROM THE GLOBAL PARAMETERS WHEN
A TARGET DOES NOT HAVE THE SAME CONDITIONS AS THE SURROUNDING
ENVIRONMENT (SEE SECTION 8-17-2 GLOBAL ENVIRONMENTAL SETTINGS).
THERMOGRAPHERS SHOULD ENSURE THEY UNDERSTAND THE PRINCIPLES
OF HEAT TRANSFER AND RADIOMETRY BEFORE BEGINNING AN INSPECTION.
SEE CHAPTER... RADIOMETRY
USERS CAN ALSO CHANGE THE REGIONAL PARAMETER SETTINGS BY
VISITING THE ENVIRONMENTAL PARAMETERS TAB OF THE SETTINGS MENU.
BE ADVISED, THE FULL FRAME SETTINGS AFFECT THE CENTER POINT
TEMPERATURE MEASUREMENTS.