IceCure Medical Ltd.
DSR3210000 rev. D
ProSense™
Confidential
User Manual
85
Return to first page
3.
When an introducer to guide the Cryoprobes to the tissue location is used, an
introducer with tight dimensions should be used. The cryoablation may be
affected by the use of an introducer that is too big and create an air barrier
between the freezing power of the probe and the tissue, inject 5cc of sterile
saline or water into the introducer lumen to improve heat exchange between the
probe and the tissue, if too big Introducer was chosen.
For optimal probe positioning please note the following:
Probe Diameter/
Gauge
Shaft Length +
Tip ±3mm
Tip Shape
Ice Ball Shape
Cool Zone Center
From needle tip
[mm]
Part Number
3.4 mm / 10G
127
mm
Trocar
Spheric
12 mm
FAP7100000
3.4 mm / 10G
140
mm
Trocar
Elliptic
20 mm
FAP7200000
3.4 mm / 10G
172 mm
Pencil
Spheric
12 mm
FAP7300000
3.4 mm / 10G
185
mm
Pencil
Elliptic
20 mm
FAP7400000
3.4 mm / 10G
185
mm
Trocar
Elliptic
20 mm
FAP7410000
3.4 mm / 10G
285
mm
Pencil
Elliptic
20 mm
FAP7910000
3.4 mm / 10G
393 mm
Pencil
Spheric
12 mm
FAP7500000
3.4 mm / 10G
403 mm
Pencil
Elliptic
20 mm
FAP7700000
2.4 mm / 13G
124
mm
Trocar
Spheric
10 mm
FAP7600000
2.4 mm / 13G
134
mm
Trocar
Elliptic
14.5 mm
FAP7800000
2.4 mm / 13G
180
mm
Trocar
Elliptic
14.5 mm
FAP7820000