Do not use a liquid nitrogen Dewar if it is damaged.
You can tell that a
Dewar is damaged if after filling it, frost appears on the outer wall of the
container. Return the Dewar to an IceCure-Medical technician or an
authorized distributor for inspection.
Warning
Before beginning a procedure on a new patient, the Dewar MUST be
properly filled and placed in the system as described in section 5.2.6.1 of this
manual.
2.4.5
Burn hazards
The Cryoprobe tip can reach very low temperatures.
Warning
Portions of the cryoprobe other than the freeze zone, including the plastic
cover that is located near the cryoprobe handle, may become cold and cause
tissue damage. If unwanted freezing occurs, stop the pre-test process by
pressing "Cancel".
Warning
Portions of the cryoprobe other than the freeze zone, including the plastic
cover that is located near the cryoprobe handle, may become cold and cause
tissue damage. If unwanted freezing occurs, immediately stop the freezing
process.
To prevent injury, cryoprobes must be closely observed during use for signs
of unwanted freezing.
2.4.6
Grounding
Warning
To avoid risk of electric shock, this equipment must only be connected to a
supply main with protective grounding.