IceCure Medical Ltd.
DSR3210000 rev. D
ProSense™
Confidential
User Manual
68
Return to first page
Ice ball measurement ranges are based on bench testing done at room temperature gel.
Isotherm measurements in soft tissue tumors may vary depending on clinical application.
Times and sizes are approximate and should not preclude clinical judgment.
2.4 Probe Spheric (ice ball shape)
(FAP7600000)
Ice Ball Dimensions at Various Temperatures over Time
-3°C*
-20°C
-40°C
Time
(minutes)
Width
(mm)
Length
(mm)
Width
(mm)
Length
(mm)
Width
(mm)
Length
(mm)
2
20
24
16
20
12
17
3
23
27
18
22
14
18
4
24
30
20
24
15
18
5
28
31
22
25
16
19
6
30
32
23
26
16
20
8
33
36
24
27
17
20
10
35
38
26
29
17
21
15
39
42
29
31
19
21
*The ice front of the ice ball visible under imaging is around -3°C.
2.4 Probe Elliptic (ice ball shape)
(FAP7800000; FAP7820000)
Ice Ball Dimensions at Various Temperatures over Time
-3°C*
-20°C
-40°C
Time
(minutes)
Width
(mm)
Length
(mm)
Width
(mm)
Length
(mm)
Width
(mm)
Length
(mm)
2
19
34
16
30
13
27
3
24
36
20
32
16
28
4
28
38
23
34
17
28
5
30
41
24
35
18
29
6
32
43
26
36
19
29
8
35
46
28
38
21
30
10
38
47
30
39
22
31
15
46
51
34
39
24
32
*The ice front of the ice ball visible under imaging is around -3°C.