Figure 81:
Pause (No 1) and Play (No 2) buttons during the freeze cycle.
The pause time does not affect the freeze preprogrammed time, but depending on when
and how long the pause time is defined, it may affect the ice ball size.
The physician should always monitor the ice ball size during the entire procedure.
IceCure Medical recommends using Pause only if necessary.
5.3.4.1
Manual freeze mode
Choosing the Manual freeze mode will open the treatment screen for manual mode as
show in Fig. below.
•
To
activate
the freeze cycle, press the Action button on the straight cryohandle
and hold it for one second or press the FREEZE icon on the screen. The
BLUE
LED
will light up around the Action button and freeze cycle will begin.
Figure 82: Manual Mode screen: press on Freeze to start the freeze cycle.