M01/M02 • Denervated 1/2 (medical programs)
Cable type: TYPE 1 (see Figure 1 - Cable type.) active channels CH1 and
CH3
Stimulation programs of the denervated muscle.
Electrodes’ positioning: use 2 big electrodes, better if wet and in a sponge,
placed at the two ends of the muscle to be treated.
−
To move the cursor and select the stimulation parameter, use the
right and left arrow keys, change the values with the up and down
arrow keys.
−
Once all the stimulation parameters have been set, position the
cursor on "OK" to memorize the program.
−
To start the stored program, press the OK button.
−
The end of the program is signaled with a triple acoustic signal.
−
With pulses greater than 100ms, set maximum frequency to 1 Hz.
Table of settable values:
Parameter Description
Minimum
Maximum
Frequency
0,25 Hertz
2 Hertz
Impulse width
40 milliseconds
400 milliseconds
Program total time
1 minute
60 minutes
ATTENTION
: In the case of programs M01 and M02, we recommend
the use of rectangular electrodes (measuring 50x90 mm) for stimulation at
medium-high intensity. With smaller electrodes, the device could generate
an alarm and, therefore, not perform the treatment correctly.
M03/M04/M05/M06 • TENS memory 3/4/5/6 (medical programs)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
Available programs to set rapid tens, endorphinic tens, maximum values
Electrodes’ positioning: forming a square above the area to be treated as
Figure 2
.