Intensity: must be such to assure good muscular contractions during
treatment, but not so strong as to generate soreness (20
÷
25mA).
Session duration: 25 minutes.
Electrodes’ positioning: from the
Positions manual:
Muscles Group all you want to stimulate in a sequence
Phase 1
Phase 2
Phase 4
3min
Warm up
20min sequential contractions on 4 channels
6s duration each
2min
Recovery
E17 • Sequential tone 2 (non-medical program)
Cable type: TYPE 1 (see Figure 1 - Cable type.)
This program produces rhythmic contractions with a stimulation frequency
typical of fast-twitch fibres. Thanks to the high stimulation frequency it is
suitable for increasing muscle strength sequentially.
The programs produce sequential phasic contractions on four channels in
phase 2 and 3. These programs can be carried out using self-adhesive
electrodes. Unlike the previous program, this one uses a higher stimulation
frequency during the contraction phase and therefore works mainly on
fast-twitch fibres.
Intensity: must be such to assure good muscular contractions during
treatment, but not so strong as to generate soreness (20
÷
40).
Session duration: 25 minutes.
Electrodes’ positioning: from the
Position manual
based on the area to be
treated.
Phase 1
Phase 2
Phase 4
3min
Warm up
20min sequential contractions on 4 channels
6s duration each
2min
Recovery
Programmable memories
Prg
Medical
Prog.
Yes/No
Description
PHASE 1
M01-
M02
Yes
Denervated
1/2
Total time from 1 to 60min
Adjustable frequency 0,25-2Hz
Adjustable pulse width 40 -400ms
M03-
M06
Yes
TENS MEM
Total time from 1 to 60min
Adjustable frequency 0,25-120Hz