IACER S.r.l.
32 di 36 MNPG20 Rev. 03 08/10/10
PRG
Hz
Impulse
width
Contraction/relaxation
time
Therapy
time
Program details
P12
1000Hz
100µs
Continuous
20 min.
Standard ionophoresis 1,
monophasic impulse
P13
1200Hz
100µs
Continuous
20 min.
Standard ionophoresis 2,
monophasic impulse
For ionophoresis program the intensity must be strong enough to produce a relevant perception, near
pain, till the muscles surrounding the area treated begin to contract.
Position of electrodes:
position
the electrode with the drug on painful area, and the other electrode on the other side.
PRG
Hz
Impulse
width
Contraction
/relax. time
Therapy
time
Program details
P14
P15
P16
2÷150Hz
(1Hz
step)
50÷300µs
(5 µs
step)
0÷30 sec
/0÷30 sec
10÷90 min
or
continuous
Free memory, adjustable slope
0÷10
sec.,
channel
output
(simultaneous, alternated, delayed
1÷6 sec.)
Free memories allow the user to adjust the parameters regarding wave frequency and impulse width,
slope and channel output (simultaneous, alternated, delayed 1÷6 sec.), therapy time. We suggest to use
free memories after consulting medician or using specified suggestion of therapist/specialist.
Electrodes and probe positions, perineal rehab