HANDEL CLEANING
The optical fiber laser handpiece is a delicate component which requires
suitable daily maintenance.
The following recommendations should be observed to protect the fiber and
lens from damage.
IACER's recommendations:
1.
use a soft cloth to remove traces of dust;
2.
clean the external parts with neutral, non-abrasive products;
3.
use a cloth to dry the external parts with care.
REPLACEING THE FUSES
In order to replace correctly the fuses on the back of the device, follow the
instructions below:
-
turn off the equipment;
-
unplug the equipment;
-
use a screwdriver to open the fuse-holder tray, making sure to insert
the screwdriver in the slot of the fuse-holder tray and lever it
outwards;
-
remove the fuse-holder structure, sliding it along the guide;
-
remove the fuses and replace them with the same number of new
ones;
-
insert the fuse-holder structure back in the tray, sliding it along the
guide;
-
close the plastic cover of the fuse-holder tray and use the tool to
fasten the fuse-holder caps.
It is advisable to have the fuses replaced by staff who have received suitable
technical training, in the interest of safety.
Power supply
230Vac 50Hz
Fuses
LA8000/LA10000
View of the back
Summary of Contents for I-TECH LA8000
Page 1: ...USER MANUAL MNPG179Rev 2of03 07 2019 Laser therapy...
Page 2: ......
Page 4: ...IACER Srl ii MNPG179...
Page 10: ...IACER Srl 6 MNPG179 Device and command description...
Page 11: ...IACER Srl 7 MNPG179 Labels...
Page 44: ......