Characteristic
Specification
Programmable treatment time
Between 1 second and 30min
Free memory
10
External dimensions (width x height x depth) 18x18x35cm
Weight of the body of the machine
2Kg
Distance between the laser emission point
and the skin
Around 38 mm
Spot area
Around 10mm
2
±3%
Operating
conditions
environmental
temperature
From +10°C to +40°C
relative humidity
From 15 to 93% without
condensation.
atmospheric pressure
500-1060 hPa
Conditions
for
storage/transport
with the device in
the
packaging
provided by the
company
environmental
temperature
From +5 to +40 °C
relative humidity
From 15 to 93% without
condensation.
atmospheric pressure
500-1060 hPa
The useful life of the device is 10 years.