IACER Srl
29
MNPG179
Troubleshooting
The I-TECH LA8000/10000 laser therapy machines were designed and
created using state-of-the-art technology and quality components to ensure
efficient and reliable performances.
Should any operational issue occur, however, the operator is recommended
to consult the guide below before contacting an authorized assistance
center.
PROBLEM
POSSIBLE CAUSE
SOLUTION
The LCD display
on the front panel
does not turn on:
the
equipment
does not work.
The plug is not inserted
properly in the mains
power socket.
Check operation of the
socket.
The power cable is not
inserted correctly in the
connector
of
the
equipment.
Insert the plug properly
in the socket and the
cable in the connector of
the equipment.
The power cable is worn
and damaged.
Replace the power cable.
The emergency switch is
turned off.
Turn on the emergency
switch.
One or more of the fuses
are
defective
or
damaged.
Replace the fuse(s) that
are missing, defective or
damaged.
The electronic control
circuit is defective.
Contact an IACER Srl
assistance center.
The LCD display
on the front panel
does not turn on.
Some components on the
electronic control board
are defective.
Contact
the
manufacturer.
Some
of
the
controls on the
front panel do not
work properly.
Defective
keys
or
buttons.
Contact
the
manufacturer.
The electronic control
circuit is defective.
The
equipment
turns
on,
but
emission of the
laser
is
not
satisfactory.
The parameters have not
been set correctly.
Check
the
correct
configuration
of
the
work parameters.
The laser sources do not
work or are worn out.
Check the activation of
emission of the laser
sources.