The following tables contain the Manufacturer’s declaration and
additional information required by IEC 60601-1-2:2014 (Fourth
Edition).
Phenomenon
Basic EMC standard
or test method
Port tested
Test level/
requirement
Radiated disturbance CISPR 11: 2015 +A1:
2016
Enclosure Group 1,
Class B
Electrostatic
Discharges (ESD)
IEC 61000-4-2:2008
Enclosure
±8 kV
Contact
±2 kV, ±4 kV,
±8 kV, ±15 kV Air
Radiated RF
Electromagnetic
Fields
IEC 61000-4-3:2006 +
A1:2007+A2:2010
Enclosure
3 V/m
80 MHz to 2.7 GHz
80%, 1 Hz AM
RF Wireless Comm.
(Refer to test report
clause 1.15)
Electric Fast
Transients and bursts IEC 61000-4-4:2012
AC Mains ±2 kV AC,
100 kHz PRR
Surges
IEC 61000-4-5:2014
AC Mains ±0.5 kV, ±1 kV
L1 to L2 (DM)
Conducted
Disturbances,
induced by RF fields
IEC 61000-4-6:2013
AC Mains
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM
bands between
0.15 MHz and
80 MHz
Power-frequency
Magnetic Field
IEC 61000-4-8:2009
Enclosure 30 A/m
EMC Table
Warranty Information
Manufacturer's Warranty
i-SENS, Inc. warrants that the CareSens N Plus BT Meter shall
be free of defects in material and workmanship in normal use
for a period of 5 years. The meter must have been subjected to
normal use.
The warranty does not cover improper handling, tampering,
use, or service of the meter. Any claim must be made within the
warranty period.
The i-SENS, Inc. will, at its discretion, repair or replace a defective
meter or meter part that is covered by this warranty. As a matter
of warranty policy, i-SENS will not reimburse the consumer’s
purchase price.
Obtaining Warranty Service
To obtain warranty service, you must return the defective meter
or meter part along with proof of purchase to your nearest
i-SENS Authorized Warranty Station.