28
www.i-sens.com
www.i-sens.com
29
Setting the Strip Expiration Date Indicator
Step 1
Entering the Expiration Date Setting
Press and hold the
▲
and
▼
buttons at the same time for 3
seconds to enter the expiration date settings. After all segments
flash across the screen, 'EP' will show up.
Step 2
Setting the Year
A number indicating the year will blink in the left
corner of the screen. Press the
▲
or
▼
button
until the correct year appears.
Press the
●
button to confirm the year and set the
month.
Note:
The strip expiration date is printed on the test strip vial.
Step 3
Setting the Month
A number indicating the month will blink at the
bottom of the screen. Press the
▲
or
▼
button
until the correct month appears. After setting,
press and hold the
●
button for 3 seconds to turn
off the meter.
Checking the System
You may check your meter and test strips using the
CareSens Control Solution (Control A and B).
The CareSens Control Solution contains a known
amount of glucose and is used to check that the
meter and the test strips are working properly.
The test strip vials have CareSens Control Solution
ranges printed on their labels. Compare the result
displayed on the meter to the CareSens Control
Solution range printed on the test strip vial.
Before using a new meter or a new vial of test strips,
you may conduct a control solution test following
the procedure on pages 30–31.
Notes:
• Use only the CareSens Control Solution (available for purchase
separately).
• Check the expiration date printed on the bottle. When you first
open a control solution bottle, record the discard date (date
opened plus three (3) months) in the space provided on the
label.
• Make sure your meter, test strips, and control solution are at
room temperature before testing. Control solution tests must
be done at room temperature 68–77 ˚F (20–25 ˚C).
• Before using the control solution, shake the bottle, discard the
first few drops and wipe the tip clean.
• Close the control solution bottle tightly and store at a
temperature of 46–86 °F (8–30 °C).
Control Soluti
EXP
ufacturer :
i-SENS, Inc.
anpo-daero 28-gil, Seocho-gu, Seoul, 06646
trol Solution Range Written on Str
ard Date :
PGA2E0580 REV