54
www.i-sens.com
www.i-sens.com
55
⑦
Repeat the same procedure for the
lancing device (step
①
to step
⑥
).
* After the pre-cleaning and disinfection procedures,
perform a control solution test to ensure the meter is
working properly. Verify that the test results are within
the range printed on the test strip vial. See pages 30-31
for control solution testing.
⑥
Allow the exterior to remain wet for 1 minute.
Note:
If any of the following deterioration signs appear
after pre-cleaning or disinfecting, please stop using the
system and contact Customer Service: 1-800-429-5001
(9 am–9 pm EST, Monday–Saturday).
At all other times, contact
your healthcare professional.
• When the inscriptions on the exterior of the meter
(or lancing device) have been removed
• When the color of the meter (or lancing device) has сhanged
or faded
• When cracks or roughness develop on the meter
(or lancing device)
• When a part of the segment on the meter display does
not flash
• When control solution test results do not fall within the
stated range on the test strip vial
Caution:
• Do not use other cleaners or disinfectants. Other chemicals
have not been validated and may damage the meter.
• Do not get fluids inside the meter through the test strip
port, data transmission port or battery compartment.
• Never immerse the meter or hold it under running water
because this will damage the meter.
Storage and Handling Caution:
• Do not expose the meter to direct sunlight or heat for an
extended period of time.
• Do not let dirt, dust, blood, or water enter into the meter’s test
strip port.