MULTIPARAMETER DIAGNOSTIC 

DEVICE

USER’S MANUAL

Summary of Contents for MULTICAREIN

Page 1: ...MULTIPARAMETER DIAGNOSTIC DEVICE USER S MANUAL...

Page 2: ...ce 7 The display 8 The measurement 8 Insertion of the data chip 10 How to obtain a drop of blood 11 Application of the blood 11 Procedure to test glucose 12 Procedure to test cholesterol 13 Procedure...

Page 3: ...ustic signal 20 Setting the alarms 21 Enabling data transfer to a computer 22 Maintenance and storage conditions 22 Care and maintenance 23 Storage and measuring conditions 24 Error messages 27 Replac...

Page 4: ...Reflectometric with MULTICAREIN cholesterol and triglycerides strips Self testing of glucose cholesterol and triglycerides could be helpful but any treatment has to be anyway agreed with the physicia...

Page 5: ...MULTICAREIN mobile app and cloud only for bluetooth MULTICAREIN Read carefully the instructions for use of the MULTICAREIN strips that you are using and read the instructions for use of the lancing de...

Page 6: ...e device c change parameters and values while setting 4 button press this button to a enter the MEMORY of the device b recall results stored in the MEMORY of the device c change parameters and values...

Page 7: ...to dispose of the used strip THE DISPLAY BACK 8 9 10 1 8 2 3 4 5 6 7 9 12 11 10 Data chip symbol 1 Average results symbol 2 Memory mode symbol 3 Settings mode symbol 4 Low battery warning 5 Result or...

Page 8: ...strips box How to insert the data chip 1 Remove the data chip from the strips box 2 Insert the data chip in the slot placed on the lateral side of the device 3 Gently press the data chip up to the co...

Page 9: ...eck if the code shown on the display matches the one printed on the label vial of the strips 5 If the data chip is not inserted it is not possible to perform the measurement 6 Analytes TGL TRIGLYCERID...

Page 10: ...etely dried Do not apply disinfectant or beauty cream before the finger pricking 2 Prepare the lancing device that is inside the MULTICAREIN meter box by inserting a sterile lancet Use a new and not d...

Page 11: ...rvice 2 Wash your hands and dry them thoroughly 3 Take out a strip from the vial and insert it in the device with the printed or coloured part facing up You will hear a click and the device will emit...

Page 12: ...on the display Press the corresponding button or to select the nutritional state The selected letter will be shown for two seconds on the display After that the blinking drop will appear on the displa...

Page 13: ...ile lancet Remove the first drop with a sterile gauze and gently press the finger to obtain the second drop Apply the blood on the white circular area of the strip coming out from the meter Wait until...

Page 14: ...The values previously measured will be saved in position 2 3 etc starting from the most recent How to recall from the memory a measured value and how to erase memorized values is described on page 17...

Page 15: ...F Use the strip immediately after removing from the vial and close the vial immediately with its cap The strips must not be used after the expiry date Write down the date of the first opening on the...

Page 16: ...s vial label that you are using Write down the date of the first opening on the vial of the control solution The control solutions must be used within 90 days from the first opening Do not use control...

Page 17: ...e average of the results AVG will be displayed 2 Hold on the button the number 1 will be displayed which refers to the most recent result When you release the button the stored result will be displaye...

Page 18: ...y pressing and releasing the power button 4 Press and release the or button to modify the date format mm dd or dd mm Confirm the wanted format by pressing and releasing the power button 5 Press and re...

Page 19: ...nt mg dL or mmol L Confirm the wanted unit of measurement by pressing and releasing the power button 10 Press and release the or button to modify the number of days you want the average of results 0 7...

Page 20: ...tting by pressing and releasing the power button 12 Press and release the or button to select first the hour and then the minutes of the first alarm It is possible to set 4 alarms The alarms are not a...

Page 21: ...a computer by means of the cable 3 Press and release the power button The message PC will appear on the display 4 Set up your computer software to receive information from the meter After the data ha...

Page 22: ...ning procedure only if blood or other material debris dust etc come into direct contact with the plastic support of the device If so wipe the device surface with a soft cloth dampened with 70 alcohol...

Page 23: ...or example in a hospital clean the device after each measurement with a 70 alcohol solution before proceeding with the next measurement STORAGE AND MEASURING CONDITIONS In order to get accurate result...

Page 24: ...ip please make sure that the strip does not remain lifted up with respect to its support The blood was applied on the strip before waiting for the drop blinking on the display Take a new strip insert...

Page 25: ...ove the wrong data chip and insert the correct one making sure that it corresponds to the test that you are performing Wrong reading of the data chip or data chip error Remove and insert again the dat...

Page 26: ...ormations The result is higher than the upper limit of reading of that parameter see also instructions for use of the strips Contact immediately your physician and repeat the test in the hospital Resu...

Page 27: ...are the right ones How to replace batteries 1 Let the battery housing slide out by pressing on the arrow printed on the battery compartment 2 Remove the batteries from the plastic holder 3 Place the n...

Page 28: ...negligence technical assistance performed by not authorized personnel accident abuse or by using strips which are different from MULTICAREIN will not be replaced Biochemical Systems International S p...

Page 29: ...one E mail Serial No Meter back Date of purchase In accordance with D L gs 196 2003 on the protection of personal data Biochemical Systems International S p A would like to inform you that all the inf...

Page 30: ...m 10 C to 40 C from 50 F to 104 F Humidity limits From 0 to 90 relative humidity non condensing Memory capacity 500 measurements with date and time Interface for data processing Connecting cable Speci...

Page 31: ...In Vitro Diagnosticum Read instruction for use Store at temperatures between Direct current CE mark Recyclable Number of pieces Manufacturer Catalogue number Do not dispose the device in the urban was...

Page 32: ...BIOCHEMICAL SYSTEMS INTERNATIONAL S p A Loc Palazzo del Pero 23 52100 Arezzo AR Italy Tel 39 0575 984164 Fax 39 0575 984238 www biosys it info biosys it MCIN MAN ENG ED2019MAR...