3
colony forming unit (CFU). Since the luminometer is detecting total
ATP, it is unknown whether the RLU result displayed by the
luminometer is due to the detection of microbial ATP, residual ATP,
or a combination of both. Therefore, a comparison cannot be drawn
between RLU values from ATP and standard plate counts (SPC); or
rather RLU does not equal CFU. True microbial contamination can
only be determined by a standard plate count method that can take
up to 48 hours for results. The luminometer is capable of providing
results in 15 seconds offering a rapid, real-time approach to
hygiene/sanitation monitoring. The luminometer is not meant to
replace microbial testing. It can be an effective tool when
incorporated into a Hazard Analysis Critical Control Point (HACCP),
Good Manufacturing Practices (GMP), or sanitation plan.
1.4
Test Devices and Accessories
Device
Cat No.
Description
UltraSnap
US2020
Surface total ATP test device
SuperSnap
SUS3000
High sensitivity total ATP test suited
for allergen prevention applications
AquaSnap Total
AQ100X
Water total ATP test device
AquaSnap Free
AQ100FX
Water Free ATP test device
Contamination
Control Kit
CCK-4
Cuvette based water testing kit
MicroSnap
Enterobacteriaceae
MS-EB
Enterobacteriaceae test
MicroSnap
Coliform Test
MS-COLIFORM Coliform test
MicroSnap E. coli
Test
MS-ECOLI
E. coli test
MicroSnap Total
MS-TOTAL
Total Viable Count test
ZymoSnap ALP
ZS
Alkaline Phosphatase test
CrossCheck
CX3000
Acid Phosphatase test
Calibration Kit
PCD4000
Positive and negative rod for
validating luminometer calibration
ATP Positive
Control Kit
CK25
25 vials of freeze-dried ATP to verify
efficacy of ATP test devices
See the Hygiena website for a complete list of test devices,
accessories, and other products.
www.hygiena.com