19 

Now  remove  the  testing  device  and  dispose  of  it.  Previous  test 

results can be viewed by pressing the   and   buttons (see section 

3.5

), or the results statistical data can be obtained by pressing the 

 button (see section 

3.6

). 

TIP

:  The luminometer can be recalibrated by removing the  testing 

device,  closing  the  lid  and  then  pressing  and  holding  down  the   

button for 1 second. The luminometer will then perform a 15-second 

zero

 measurement. 

4.2

 

Performing a Retest

 

 

When  viewing  a  failed    test  result  (see  sections 

3.5

 and 

4.2

), the program location can be selected 

for retesting by pressing the   button, followed by 

the   button: 

NOTE

:  It  is  only  possible  to  select  a  program 

location for retest if the most recent test result for that location was 

a failure  . 
Having  selected  retest  mode  (signified  by  the 

  icon  next  to  the 

Prog:

, perform the sample measurement in the usual way, using a 

new testing device. (see section 

4.1

). 

Once  the  retest  measurement  is  complete,  the  original  failed  test 
result is marked with the 

 icon, and the new test result is tagged 

with the   icon: 

 

 

NOTE

:  Retest  mode  is  automatically  deselected  once  the  new 

measurement is complete, or can be manually cancelled by pressing 

the   button and selecting a new program location. 

5.

 

Installing SureTrend Data Analysis Software 

The  test  results  stored  in  the  luminometer's  memory  can  be 

uploaded  to  a  PC  using  the  SureTrend  Data Analysis software. The 

software is supplied on the CD with the luminometer. 

5.1

 

Conditions of Use 

As Hygiena LLC has no control over the specification, state or use of 

any computer equipment on which this software is installed or used, 

the following Conditions of Use apply. 

Summary of Contents for EnSURE v2

Page 1: ...I Ultrasnap SureTrend and dataSURE II are trademarks of Hygiena LLC Operating Precautions and Limitations of Use The following symbol is used in this manual Description CAUTION WARNING IMPORTANT Hygie...

Page 2: ...P PROGRAM TEST POINTS AND PASS FAIL LIMITS _____________________ 10 3 3 TEST PLANS ______________________________________________________ 11 3 4 ESTABLISHING ATP PASS FAIL LIMITS _____________________...

Page 3: ...__ 23 7 TROUBLESHOOTING ________________________________________________ 24 7 1 LUMINOMETER BEEPS _______________________________________________ 24 7 2 TROUBLESHOOTING TIPS __________________________...

Page 4: ...evels of contamination on surfaces water and product samples 1 Remove swab from test tube and collect sample Refer to instructions included with testing device 2 Place swab back in test tube 3 Break S...

Page 5: ...ased from inside these cells along with any free ATP picked up from the surface by the swab is now available to react with the chemistry in the bulb of the device When the Luciferin Luciferase enzymes...

Page 6: ...ng Practices GMP or sanitation plan 1 4 Test Devices and Accessories Device Cat No Description UltraSnap US2020 Surface total ATP test device SuperSnap SUS3000 High sensitivity total ATP test suited f...

Page 7: ...lect and Exit button Up increase button Start accept button Down decrease button Program select button Statistics shortcut button The functions of the keypad are explained in detail in the following s...

Page 8: ...Cell Voltage Relative Capacity Alkaline 1 5V 1 0 NiMH 1 2V 0 6 NiCD 1 2V 0 5 The unit requires two batteries of size AA LR6 or E91 TIP For best results always use a quality brand of Alkaline battery...

Page 9: ...ely The unit will then perform an internal 15 second calibration see section 2 7 NOTE If the clock is not set the unit will automatically enter the time and date set up mode see section 3 8 before per...

Page 10: ...ement reading in Relative Light Units RLU Retested icon identifies the failed test result as having been retested see section 4 2 Retest result icon identifies the Program location or test result as a...

Page 11: ...and is flat TIP The button can be pressed at any time to access the set up menu options see section 3 When the calibration is complete the luminometer is ready to perform a sample measurement For deta...

Page 12: ...button Use the and buttons to scroll through the menu list and then press the button to select the required menu option or the button the exit the menu The following 11 menu options are available User...

Page 13: ...uminometer or set using the SureTrend software If a location name is entered and threshold set in SureTrend and then downloaded to the luminometer the threshold values can t be changed via the luminom...

Page 14: ...ocations specified in the order in which they would normally be tested Having selected the TEST PLAN menu option use the and buttons to scroll through the list of Test Plans and then press the button...

Page 15: ...guide included with your system or visit the resources available on www hygiena com 3 5 Viewing Stored Test Results The luminometer can store a maximum of 2000 test results in its internal memory eve...

Page 16: ...to exit the option 3 8 Setting the Clock Time and Date To set or change the clock time date and format select the CLOCK menu option by pressing the button Scroll to CLOCK and press Then use the and bu...

Page 17: ...ready to perform readings Quick Boot can be turned off and the luminometer will perform a more in depth system check the next time it is turned on The more in depth system check takes 60 seconds NOTE...

Page 18: ...able or disable the warning beeps This will toggle the bell icon from enabled to disabled Press the button to accept the settings increase or decrease duration 5 to 60 set duration enable or disable w...

Page 19: ...corded NOTE The Save Tests mode will be set back to ON when the luminometer is turned off and back on again When Save Tests mode is set to OFF RESULTS NOT SAVED will appear at the bottom of the LCD To...

Page 20: ...y shows the following details Various keypad options are now available as detailed in each of the following sections Button Action Section Access set up menu options 3 Select different User ID 3 1 Sel...

Page 21: ...e displayed WARNING Always ensure that the exterior of the testing device is clean and dry before inserting it into the luminometer Never insert anything other than a Hygiena testing device into the l...

Page 22: ...location was a failure Having selected retest mode signified by the icon next to the Prog perform the sample measurement in the usual way using a new testing device see section 4 1 Once the retest mea...

Page 23: ...on your computer has been disabled Open My Computer right click on the CD drive and select AutoPlay If this does not start the install program open the CD folder and run the Setup application NOTE Yo...

Page 24: ...ardware Wizard does not start check the following a You have sufficient permissions to run the wizard b Verify there are no existing hardware problems in device manager c Verify the Universal Plug and...

Page 25: ...intenance The luminometer does not require any specific routine operator or service engineer maintenance 6 1 Cleaning the Luminometer Case Clean the luminometer case when required using a dry or sligh...

Page 26: ...ocket using water or a very mild detergent solution ensuring that the pocket is completely dry and clean before putting it back into the luminometer WARNING Do not use solvents or other strong chemica...

Page 27: ...use the luminometer emits a variety of different beeping sounds Beep Type Possible Causes Short high pitched tone Luminometer turned on or off Sample measurement started Communications established wit...

Page 28: ...window is dirty Display window is scratched or dented Display or luminometer damaged or faulty Keypad button has no effect when pressed Some buttons only work when selecting particular luminometer fu...

Page 29: ...dirty or severely scratched Luminometer damaged or faulty E4 User settings undefined The user configurable settings are undefined check and re set the CLOCK format User Plan and Prog data E5 Program...

Page 30: ...n of any part of the luminometer or the use or attachment of any peripheral not supplied or specified by Hygiena will void this Warranty WARNING Use only the accessories and consumables supplied by Hy...

Page 31: ...t resolution 1 RLU Measurement time 15 seconds Measurement noise 5 or 5 RLUs User IDs 200 IDs of 16 characters Programmable result thresholds 5001 programs Programmable location names 5001 names of 20...

Page 32: ...alkaline batteries or rechargeable NiMH NiCD batteries of types specified in section 10 WARNING Do not use batteries with individual cell voltages greater than 1 65V as this will cause permanent damag...

Page 33: ...1 The luminometer is manufactured under ISO 9001 controls Declaration of Conformity The luminometer has been designed in accordance with and satisfies the requirements of article 11 of the Low Voltage...

Reviews: