19
Now remove the testing device and dispose of it. Previous test
results can be viewed by pressing the and buttons (see section
3.5
), or the results statistical data can be obtained by pressing the
button (see section
3.6
).
TIP
: The luminometer can be recalibrated by removing the testing
device, closing the lid and then pressing and holding down the
button for 1 second. The luminometer will then perform a 15-second
zero
measurement.
4.2
Performing a Retest
When viewing a failed test result (see sections
3.5
and
4.2
), the program location can be selected
for retesting by pressing the button, followed by
the button:
NOTE
: It is only possible to select a program
location for retest if the most recent test result for that location was
a failure .
Having selected retest mode (signified by the
icon next to the
Prog:
, perform the sample measurement in the usual way, using a
new testing device. (see section
4.1
).
Once the retest measurement is complete, the original failed test
result is marked with the
icon, and the new test result is tagged
with the icon:
NOTE
: Retest mode is automatically deselected once the new
measurement is complete, or can be manually cancelled by pressing
the button and selecting a new program location.
5.
Installing SureTrend Data Analysis Software
The test results stored in the luminometer's memory can be
uploaded to a PC using the SureTrend Data Analysis software. The
software is supplied on the CD with the luminometer.
5.1
Conditions of Use
As Hygiena LLC has no control over the specification, state or use of
any computer equipment on which this software is installed or used,
the following Conditions of Use apply.