background image

1.1

 

What is ATP? 

Adenosine  Triphosphate  (ATP)  is  an 

organic  molecule  that  is  used  by  living 

cells  as  the  main  source  of  energy. 

Animal, plant, bacterial, yeast, and mold 

cells  produce  and  break  down  ATP  in 

order  to  drive  a  number  of  biological 

processes. It is the main energy-transfer 

molecule in the cell. The presence of ATP on a surface indicates that 

the  surface  has  not been adequately cleaned and has the potential 

to  harbor  and  support  bacterial  growth.  The  residue  may  also 

contain  hazardous  material  including potential allergens. Therefore, 

ATP is an ideal indicator of surface cleanliness. 

1.2

 

How Does the Luminometer Work? 

The luminometer detects ATP using a chemical reaction. 

Luciferin/Luci ATP = Light 

A sample is collected using a Hygiena ATP device (see section 

1.4

). 

Surface  tests  come  with  a  buffer  moistened  swab  bud  that  aids  in 

the removal of any biological material (ATP) on either a wet or dry 

surface. Once the sample is collected and exposed to the chemistry 

(buffer)  in  the  Hygiena  device,  the  ATP  contained  inside  the 

collected cells will be released. ATP released from inside these cells, 

along with any free ATP picked up from the surface by the swab, is 

now available to react with the chemistry in the bulb of the device. 

When the Luciferin/Luciferase enzymes are exposed to ATP, light is 

produced. The light is detected and measured by the  luminometer. 

The amount of light detected is directly proportional to the quantity 

of  ATP  collected  and  accordingly  is  proportional  to  the  amount  of 

residue in the tested sample. The higher the reading, the higher the 

amount of contamination in the sample. 

1.3

 

What is an RLU? 

The luminometer displays results in Relative Light Unit (RLU) values. 

The light produced from the Luciferin/Luciferase and ATP reaction in 

the  swab  is  emitted  in  the  form  of  photons.  A  photon  is  an 

elementary  particle  and  the  basic  unit  of  light.  The  luminometer 

detects these photons and displays them directly as RLU values. The 

more light detected by the luminometer, the greater the RLU value. 

The  quantitative  RLU  reading  is  then  compared  against  user 

programmable  thresholds  to  provide  an  overall  qualitative  pass  , 

caution   or fail   result. 
The  luminometer  detects  total  ATP,  not  just  ATP  from  bacteria, 

yeast,  and  mold  but  also  the  ATP  from  anything  organic  in  the 

sample.  Therefore,  an  RLU  value  is  not  the  same  as  a  microbial 

Summary of Contents for EnSURE v2

Page 1: ...I Ultrasnap SureTrend and dataSURE II are trademarks of Hygiena LLC Operating Precautions and Limitations of Use The following symbol is used in this manual Description CAUTION WARNING IMPORTANT Hygie...

Page 2: ...P PROGRAM TEST POINTS AND PASS FAIL LIMITS _____________________ 10 3 3 TEST PLANS ______________________________________________________ 11 3 4 ESTABLISHING ATP PASS FAIL LIMITS _____________________...

Page 3: ...__ 23 7 TROUBLESHOOTING ________________________________________________ 24 7 1 LUMINOMETER BEEPS _______________________________________________ 24 7 2 TROUBLESHOOTING TIPS __________________________...

Page 4: ...evels of contamination on surfaces water and product samples 1 Remove swab from test tube and collect sample Refer to instructions included with testing device 2 Place swab back in test tube 3 Break S...

Page 5: ...ased from inside these cells along with any free ATP picked up from the surface by the swab is now available to react with the chemistry in the bulb of the device When the Luciferin Luciferase enzymes...

Page 6: ...ng Practices GMP or sanitation plan 1 4 Test Devices and Accessories Device Cat No Description UltraSnap US2020 Surface total ATP test device SuperSnap SUS3000 High sensitivity total ATP test suited f...

Page 7: ...lect and Exit button Up increase button Start accept button Down decrease button Program select button Statistics shortcut button The functions of the keypad are explained in detail in the following s...

Page 8: ...Cell Voltage Relative Capacity Alkaline 1 5V 1 0 NiMH 1 2V 0 6 NiCD 1 2V 0 5 The unit requires two batteries of size AA LR6 or E91 TIP For best results always use a quality brand of Alkaline battery...

Page 9: ...ely The unit will then perform an internal 15 second calibration see section 2 7 NOTE If the clock is not set the unit will automatically enter the time and date set up mode see section 3 8 before per...

Page 10: ...ement reading in Relative Light Units RLU Retested icon identifies the failed test result as having been retested see section 4 2 Retest result icon identifies the Program location or test result as a...

Page 11: ...and is flat TIP The button can be pressed at any time to access the set up menu options see section 3 When the calibration is complete the luminometer is ready to perform a sample measurement For deta...

Page 12: ...button Use the and buttons to scroll through the menu list and then press the button to select the required menu option or the button the exit the menu The following 11 menu options are available User...

Page 13: ...uminometer or set using the SureTrend software If a location name is entered and threshold set in SureTrend and then downloaded to the luminometer the threshold values can t be changed via the luminom...

Page 14: ...ocations specified in the order in which they would normally be tested Having selected the TEST PLAN menu option use the and buttons to scroll through the list of Test Plans and then press the button...

Page 15: ...guide included with your system or visit the resources available on www hygiena com 3 5 Viewing Stored Test Results The luminometer can store a maximum of 2000 test results in its internal memory eve...

Page 16: ...to exit the option 3 8 Setting the Clock Time and Date To set or change the clock time date and format select the CLOCK menu option by pressing the button Scroll to CLOCK and press Then use the and bu...

Page 17: ...ready to perform readings Quick Boot can be turned off and the luminometer will perform a more in depth system check the next time it is turned on The more in depth system check takes 60 seconds NOTE...

Page 18: ...able or disable the warning beeps This will toggle the bell icon from enabled to disabled Press the button to accept the settings increase or decrease duration 5 to 60 set duration enable or disable w...

Page 19: ...corded NOTE The Save Tests mode will be set back to ON when the luminometer is turned off and back on again When Save Tests mode is set to OFF RESULTS NOT SAVED will appear at the bottom of the LCD To...

Page 20: ...y shows the following details Various keypad options are now available as detailed in each of the following sections Button Action Section Access set up menu options 3 Select different User ID 3 1 Sel...

Page 21: ...e displayed WARNING Always ensure that the exterior of the testing device is clean and dry before inserting it into the luminometer Never insert anything other than a Hygiena testing device into the l...

Page 22: ...location was a failure Having selected retest mode signified by the icon next to the Prog perform the sample measurement in the usual way using a new testing device see section 4 1 Once the retest mea...

Page 23: ...on your computer has been disabled Open My Computer right click on the CD drive and select AutoPlay If this does not start the install program open the CD folder and run the Setup application NOTE Yo...

Page 24: ...ardware Wizard does not start check the following a You have sufficient permissions to run the wizard b Verify there are no existing hardware problems in device manager c Verify the Universal Plug and...

Page 25: ...intenance The luminometer does not require any specific routine operator or service engineer maintenance 6 1 Cleaning the Luminometer Case Clean the luminometer case when required using a dry or sligh...

Page 26: ...ocket using water or a very mild detergent solution ensuring that the pocket is completely dry and clean before putting it back into the luminometer WARNING Do not use solvents or other strong chemica...

Page 27: ...use the luminometer emits a variety of different beeping sounds Beep Type Possible Causes Short high pitched tone Luminometer turned on or off Sample measurement started Communications established wit...

Page 28: ...window is dirty Display window is scratched or dented Display or luminometer damaged or faulty Keypad button has no effect when pressed Some buttons only work when selecting particular luminometer fu...

Page 29: ...dirty or severely scratched Luminometer damaged or faulty E4 User settings undefined The user configurable settings are undefined check and re set the CLOCK format User Plan and Prog data E5 Program...

Page 30: ...n of any part of the luminometer or the use or attachment of any peripheral not supplied or specified by Hygiena will void this Warranty WARNING Use only the accessories and consumables supplied by Hy...

Page 31: ...t resolution 1 RLU Measurement time 15 seconds Measurement noise 5 or 5 RLUs User IDs 200 IDs of 16 characters Programmable result thresholds 5001 programs Programmable location names 5001 names of 20...

Page 32: ...alkaline batteries or rechargeable NiMH NiCD batteries of types specified in section 10 WARNING Do not use batteries with individual cell voltages greater than 1 65V as this will cause permanent damag...

Page 33: ...1 The luminometer is manufactured under ISO 9001 controls Declaration of Conformity The luminometer has been designed in accordance with and satisfies the requirements of article 11 of the Low Voltage...

Reviews: