Preparing the CTU for operation
CTU 2xxx series
en-US
Page 60 / 108
BeWa CTU2000 3160738h en-us 2018-11-15.docx
2018-11-15
Clean-room conditions in the analysis chamber - Rinsing of the inner
chamber
Before starting a series of contamination tests, cleanse the clean room of the
CTU in a defined manner using the volume or time/duration in the rinsing
program "Cleansing the clean room automatically".
During operation, as well as between extractions, make sure that the clean
room is always closed and is opened for only a short time for the insertion of
the component to be analyzed. Only in this way can you ensure that
contamination from the surrounding area is as minimal as possible.
Performing the spray extraction method
CAUTION
Spraying "G60 Spezial" test fluid
Health hazard
►
Keep the door to the analysis chamber closed
as far as possible during the spray extraction
process.
►
Always wear a face shield.
►
Always wear protective gloves.
Manual spray extraction enables you to rinse components in a defined
manner as described in the "Functional description" chapter.