Operating the text display TD 200
CTU 2xxx series
en-US
Page 54 / 108
BeWa CTU2000 3160738h en-us 2018-11-15.docx
2018-11-15
Performing the contamination test
The "Contamination Test" operating mode can be used to rinse components
in a defined manner.
CAUTION
Spraying "G60 Spezial" test fluid
Health hazard
►
Keep the door to the analysis chamber closed
as far as possible during the spray extraction
process.
►
Always wear a face shield.
►
Always wear protective gloves.
After switching on the CTU and switching to "Operation" mode, the following
message appears on the display:
Pressing the
button takes you
to the main menu.
Press the
button and
.
Press the
/
buttons to
scroll through the 10 rinsing volume
programs. Confirm your selection
with the
button.
Select the volume value by pressing
the
button or change the
value by pressing the
button.
The value will flash and can be
changed by pressing the
Contamination Test
Press OK button to start
Select rins. 0-9
Rinsing volume no. 0
Rinsing volume no.: 0
Rinsing volume no.: 1
Volume = 0.1 L
OK or up/down