Operating the text display TD 200
CTU 2xxx series
en-US
Page 57 / 108
BeWa CTU2000 3160738h en-us 2018-11-15.docx
2018-11-15
before the set cleaning time elapses.
There is still analysis fluid in the
collecting pan after the cleaning time
has elapsed.
Only press the
button
once the clean room has been
completely emptied.
Automatic reservoir change-over
During the contamination test or inner chamber cleaning, the fill levels of
reservoirs
B1
and
B2
are continuously monitored.
When the lower fill level is reached in one reservoir, the system automatically
changes over to the other reservoir.
This change-over takes some seconds and is indicated by the signal lamp.
-
The "CHANGE-OVER reservoir" signal lamp flashes until the remaining
volume has been conveyed.
-
The test fluid supply is stopped.
-
The "CHANGE-OVER reservoir" signal lamp lights up until the change-
over is completed.
-
If the contamination test is completed before the remaining volume has
been consumed, the system will change over as soon as the
contamination test is finished.
Cleaning completed
Press OK to continue