8.1.3.
Pattern Calibration
Procedures for pattern calibration
①
Press MENU button and select CALIB.
②
Press FRAME button and select PATTERN.
③
Press BOTH button.
④
Load the Standard Pattern.
⑤
Press BOTH button to start calibration.
Next, follow the instructions on displayed.
NOTE
The result of Automatic Calibration will be automatically saved.
Be sure to check the related information before executing the calibration work because
you cannot return to the previous state.
Press STOP button to stop the calibration.