HPE-810
HUVITZ
8.
System Maintenance
8.1.
Automatic Calibration of Frame Reader
Automatic Calibration of Frame Reader is required with the following cases,
When the Size or Axis value of the traced data seems to have a problem,
When the Frame Reader seems to have a problem in operating normally.
Automatic calibration of frame reader consists of three parts: Stroke, Frame, Pattern. If any changes
in Stylus occur, Stroke calibration must be executed at first. The order of calibration is like this:
Stroke -> Frame -> Pattern.
8.1.1.
Stroke Calibration
Procedures for stroke calibration
①
Press MENU button and select CALIB.
②
Press FRAME button and select STROKE.
③
Press BOTH button.
④
Pull the Gripping bars to the maximum.
⑤
Press BOTH button to start calibration.
Next, follow the instructions displayed.
8.1.2.
Frame Calibration
Procedures for frame calibration
①
Press MENU button and select CALIB.
②
Press FRAME button and select FRAME.
③
Press BOTH button.
④
Load the Standard Frame.
⑤
Press BOTH button to start calibration.
Next, follow the instructions displayed.