10
no apparent effect, so you can press “Enter” key in the current room temperature to test.
②
Boosting----On reaching the required test temperature, or after pressing “Enter” key in the
heating process, it is automatically transferred to boost state, with voltage value
increasing and adjustment. When the voltage reaches a preset voltage value, or in the
process of boosting press "Enter" key, the instrument will stop boosting, and enter the
test status.
③
Calibration results----After dry cell calibration is finished, the apparatus automatically
displays the calibration results, as shown in Fig.7.
Fig.7
5
、
Automatic measurement
Remove the inner electrode from the outer electrode, and then take measured oil
samples 40ml and inject into the outer electrode of test cell, before entering the automatic
measurement. Align the locating slot of inner electrode with the positioning key of outer
electrode and put the inner electrode into the outer electrode. And also connect the test cable
and temperature probe cable.
Press “Automatic measurement” key, and go to the auto-measuring screen in Fig.8.