Press the "select" button, move the cursor to the instrument calibration, automatically
select "Yes" and press "Enter" button, enter the interface shown in Fig.15.
Fig.15
Correctly connect the standard. As shown in Fig.15, if you want to verify dielectric loss
factor, please click the “LOSS” button. If checking the resistance value, please press the
“RES” button. Press “Back” button to return to the main menu.
In the calibration process, the instrument can be operated with lid opening, but we must
be ware of electric shock.
The screen automatically displays the validation results, as shown in Fig.16, 17. Record
the check results, and convert the gear of the standard, then press the “dielectric loss” key or
the “resistance” button, and continue to the verification of the standard’s next gear.