QUICKLAN 6050N
EN - 7
5.1.3. Setting the cable type
1. Switch on the instrument by pressing the key.
2. Press and hold the
SET/UNIT
key for 3s. The screen in Fig. 3
–
left side appears on
the display.
3. Press the arrow keys
or
until you display the screen in Fig. 5 - left side
Fig. 5: Setting the cable type
4. Press the
SET/UNIT
key. Parameter
“
0
”
flashes on the display. This parameter allows
selecting one of the 10 numeric configurations saved by the user according to the type
and possible calibration of cable length (see § 5.5.1).
5. Press the arrow key
or
to select the value among the options
0 ÷ 9
.
6. Press the
SET/UNIT
key to go to the selection of cable type. Parameter
“
CA5
”
flashes
on the display.
7. Press the arrow key
or
to select the cable type among the options:
CA5
(CAT5), CA6 (CAT6), AJII (RJ11), COA (COAX)
8. Press the
SET/UNIT
key to go to the selection of cable type with RJ45 connector.
Parameter
“
StP
”
flashes on the display.
9. Press the arrow keys
or
to select the possible options.
StP
(shielded STP
cable) or
UtP
(unshielded UTP cable)
10. Press the
TEST/
key to save all settings and go back to main screen.
5.1.4. Setting cable length measuring unit
1. Press and hold the
L/W
key and switch on the instrument by pressing the
key. The
instrument shows the screen in Fig. 6
–
left side.
Fig. 6: Setting length measuring unit
2. Press the arrow keys
or
to select measuring unit
“
Mt
”
(meter) or
“
Ft
”
(foot)
flashing on the display.
3. Press the
TEST/
key to save setting and go back to main screen.