I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
EN - 63
8.2. CONNECTING BY WIFI
CAUTION
The connection between instrument and PC is realized by means of WiFi
interface which is normally active on the instrument. The WiFi function is not
available if in the initial screen of the instrument (see § 4.4) the “WiFi-OFF”
message is displayed
In order to transfer the data onto a PC it is necessary to install the
management software Topview in advance, and WiFi interface shall be
available on the PC itself
Before connecting it is necessary to enable WiFi connection on the
instrument as below mentioned, and select/connect the “wireless network”
(WiFi) which the instrument makes available
Before connecting, it is necessary to select the “WiFi” port on the “PC
Instrument Connection” of
TopView
software and refer to the program on-
line help.
To transfer the saved data to the PC, keep to the following procedure:
1. Switch on the instrument by pressing
ON/OFF
key
2. Press
ESC/MENU
key to open the main menu.
3. By means of arrow keys (
,
) select “
PC Connection
” to access the PC mode
section and confirm by
ENTER
15/05/10 15:34:26
I - V
I-V Test
EFF Yield Test
SET
Settings
DB
Modules
MEM
Data Recall
PC
PC Connection
E N T E R t o s e l e c t .
M E N U
4. The below screen is displayed by the meter:
15/05/10 15:34:26
PC – RS232
WiFI ON
M E N U
5. Enable the WiFi connection on the PC (if necessary, it is possible to use a WiFi
adaptor). Select and connect the WiFi network which the instrument makes available in
“Network connection
Wireless networks” section
6. Use data management software TopView to transfer the content of the instrument
memory to the PC (see software on-line help for all details)