I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
EN - 34
9. Press
ENTER
key, select the “
Settings
” item and confirm
again with
ENTER
to access the next screen which allows
setting the type of PV module and the number of modules
relative to the PV string under test.
15/05/10 15:34:26
V d c
=
0 . 0
V
I r r
= - - - W / m 2
T c
= - - - ° C
Module: PANEL01
S e t t i n g s
M e a s . T y p e
Select
I - V
10. By means of arrow keys (
,
) select the type of PV
module included in the database of meter (§ 5.3.1).
11. By means of arrow keys (
,
) select the item “N. of
Mod.” and by means of arrow keys (
,
) set the number
of modules of the PV string under test. The maximum
number of modules is
50
12. By means of arrow keys (
,
) select the “Oper. yrs” item
and by means of arrow keys (
,
) set the service operating
years of the PV module/string/field from start installation
(see § 6.2.3). The maximum set value is
25.0 (0.5 = 6
months)
15/05/10 15:34:26
Type :
SUNPOWER 210
N . o f M o d . : 1 5
O p e r . y r s : 4 . 5
T e m p : A u t o
S t a r t
: M a n u a l
P m a x =
2 1 0
V o c =
4 7 . 7 0
V m p p = 4 0 . 0 0
I s c = 5 . 7 5
I m p p = 5 . 2 5
S E T
13. By means of arrow keys (
,
) select the type of PV module included in the database
of meter (§ 5.3.1). By means of arrow keys (
,
) select the “Temp” item and by
means of arrow keys (
,
) choose the type of module temperature among the modes
listed:
Auto
”
automatic measurement performed by the meter depending on the
measured value of open voltage of the modules
(recommended method)
“
Manual
”
setting of a module temperature value by the operator on the
corresponding “
Value
” field
“
Aux
”
setting of temperature of module performed by auxiliary probe
14. By means of arrow keys (
,
) select the “Start” item and by means of arrow keys
(
,
) choose the type of measurement activation between the modes listed:
Auto
”
test automatically activated by the instrument with stable values of voltage
among inputs for at least 1s and with values within the measurement range (see §
10.2)
“
Manual
”
test activated from the user by press of
GO/STOP
key
15. Confirm with
SAVE
key or press
ESC/MENU
key to exit without saving
16. Mount the M304 (inclinometer) supplied accessory, place and hold it on the plane of module.
Verify that the sun shadow falls on the disc within the “limit internal circle” on the
disc itself (see Fig. 7b). If this is not the case, the incidence angle between sun rays
and the module surface is too high and not complying with the test conditions
declared by the module manufacturer. As a consequence the measurements
performed by the meter are not ok and all measurements steps shall be repeated at
another daytime
17. Fix the support of the cell to the module by using the supplied set of screws and mount the
reference cell on it possibly with
output terminals downwards with respect to the
module
. Rotate the cell upwards to lean it on the small wing of the support so as the cell is
perfectly parallel to the plane of module, then fix it with the relevant screws supplied
18. Connect the output of reference cell (depending on the type of module under test) of
reference cell at the
IRR.
input of meter by using the cable supplied with the cell