FULLTEST
3
EN - 27
RAMP 50% mode:
Figure 16: Presence of test voltage with regard to pressed START button and set time in RAMP 50%
mode
1)
Select DIELECTRIC measurement by pressing the
FUNC
hard key first.
2)
Check selected mode and correct it if needed by pressing the
MODE
(5) touch-screen key first.
MANUAL, RAMP or BURN mode can be selected.
3)
Check selected test voltage (250 up to 5100V) and correct it if needed by pressing the
U
TES NOM
(6) touch-screen key first.
4)
Check selected limit current and correct it if needed by pressing the
LIMIT
(7) touch-screen key
first. Four independent preset limit currents are available for quicker operations. Select the one
closest to wished value and modify it by using the
+
and
―
touch screen keys if needed.
5)
Check selected character of displayed current (I
APP
or I
REAL
) and correct it if needed by pressing
the
CHAR
(8) touch-screen key first.
6)
Select measurement screen by pressing the
(4) touch-screen key and check all settings again.
7)
Connect the test leads according to the figure below.
Figure 17: Connection of test leads
8)
Carry out the test by pressing the
START
/STOP button. Warning with explanation of how to
connect test leads with regard to selected test voltage will be displayed. Check the connection then
confirm it by pressing the
YES
touch-screen key, the message “READY” will appear and will stay
present for 10 seconds.
START
/STOP button is active while “READY” message is present. Press
and keep pressing the
START
/STOP
button, test voltage will be applied to test terminals.