COMBI521
EN - 91
4. The screen on the side shows the values of the electric
quantities in real time. Symbols “
i
” and “
c
” respectively
indicate the inductive or capacitive nature of the load.
PQA
15/10 – 18:04
V L - N = 2 3 0 . 5 V
I = 2 7 . 3 A
P = 5 . 9 1 k W
Q = 2 . 1 5 k V a r
S = 6 . 2 9 k V A
P f = 0 . 9 4 i
C o s
= 0 . 9 4 i
Par 100A
MODE
FS
5. Press the
SAVE
key to store the test result in the instrument’s memory (see §
7.1
) or
the
ESC/MENU
key to exit the screen without saving and go back to the main menu.
6.
Use the arrow keys
(
,
)
in the
MODE
function to select
the option “
ArmV
” (voltage harmonics) or “
ArmI
” (current
harmonics). Subsequently, the instrument shows a screen
similar to the one on the side, in which the considered
amplitude of harmonics is always indicated in a
percentage
value with respect to the fundamental.
The histogram graphic of the percentage amplitudes of
the fundamental and of the voltage VL-N or current
harmonics with a
DC value up to the 25th
(the value of
the fundamental H01 is always considered equal to 100%
and is not shown) further to the value of THD% (see §
12.12) are shown on the display. The following controls
are available:
Pag
allows changing the page where harmonics are
displayed
Hxx
allows moving the cursor to increase/decrease
the harmonic order within the page
PQA
15/10 – 18:04
H 0 2 = 1 0 . 0 % T H D V = 1 . 8 %
Harm V
100A
MODE
FS Pag
Hxx
7. Press the
SAVE
key to store the test result in the instrument’s memory (see §
7.1
) or
the
ESC/MENU
key to exit the screen without saving and go back to the main menu