Calibrating
the
EMI
Receiver
When
either
the
HP
8542E
or
HP
8546A
is
turned
on,
the
receiver's
amplitude
calibration
data
for
INPUT
1
and
INPUT
2
is
veried.
If
the
data
is
valid,
it
is
downloaded
to
the
RF
lter
section
from
the
nonvolatile
storage
in
the
receiver
RF
section.
If
the
data
is
not
valid,
the
RF
lter
section
will
use
its
standalone
factory
calibration
data
and
display
a
\Cal
Needed "
message
.
Note
PRESET
or
IP
does
not
download
the
receiver's
calibration
data.
Using
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
FETCH
T
o
retrieve
the
receiver's
calibration
data
from
internal
memory
,
press:
4
CALIBRA
TE
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
FETCH
If
the
data
is
not
valid
when
the
calibration
data
is
retrieved
from
memory
using
the
cal
fetch
function,
one
or
both
of
the
following
messages
will
be
displayed:
INPUT
1:
Data
Not
valid
CAL
INPUT
1
required
INPUT
2:
Data
Not
valid
CAL
INPUT
2
required
V
erifying
the
Receiver's
Calibration
1.
T
o
verify
the
amplitude
calibration
data,
press:
4
CAL
CHECK
5
2.
If
the
data
is
not
valid,
the
calibration
check
procedure
is
stopped
and
the
message
\Cal
Needed "
is
displayed.
3.
If
the
data
is
valid,
the
calibration
check
is
completed
and
the
message
\Calibration
OK "
is
displayed.
Note
Only
the
amplitude
accuracy
at
300
MHz
62
dB
is
veried
when
a
4
CAL
CHECK
5
is
executed.
T
o
ensure
the
receiver
meets
its
specications
at
all
frequencies
,
a
calibration
must
be
performed.
Calibration
2-3